Published ISO/IEC 17025:2017 scope of accreditation
FCI's accredited calibration and measurement capability (CMC), grouped by discipline. Ranges and expanded uncertainties are reported at a 95% confidence level (k=2). Accredited by IAS under certificate CL-112.
Dimensional
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| Micrometers (direct) | 0.0625 in – 60 in | (70 + 7.5L) µin | Gage blocks + micrometer standards |
| Calipers (0–1 in) | 0 – 1 in | 280 µin | Direct measurement, gage blocks |
| Calipers (1–60 in) | 1 in – 60 in | 320 µin | Direct measurement, gage blocks |
Dimensional measurements are traceable to the SI metre through NIST-traceable gage-block and micrometer standards. Uncertainties are evaluated per the ISO/IEC Guide 98-3 (GUM) and reported at a 95% confidence level (k=2) on every ISO/IEC 17025:2017 certificate.
Mechanical
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| Speed (non-contact, generate)⁴ | 6 – 99,999 rpm | 0.011 – 0.59 rpm | HP 3325B |
| Scales & balances (linearity) | 200 mg – 15 kg | 0.070 mg – 130 mg | Class 1 / Class 3 weights |
| Scales & balances (lb) | 33 lb – 400 lb | 0.0058 – 0.055 lb | Class 7 weights |
| Force gauges | 0.5 – 10 lbf | 0.0041 – 0.054 lbf | Class 7 weights |
| Force (measure)³ | 0 – 225,000 lbf | 18 lbf – 1.6e3 lbf | Interface 1211NE-10K-B · VLC-120 · Tetrahedron 300K |
| Hand torque tools (CW + CCW) | 10 – 3,120 lbf·in | per ISO 6789-1 | Larson UTWCS (ISO 6789-1) |
Speed, mass, force, and torque results are traceable to NIST through reference frequency, certified weight sets, reference load cells, and torque standards. All uncertainties are reported at a 95% confidence level (k=2) on each ISO/IEC 17025:2017 certificate.
Thermal / Environmental
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| Simulated temperature (thermocouple) | −250 °C – 1820 °C | 0.08 – 0.73 °C | Fluke 7526A (direct) |
| Simulated temperature (RTD) | −200 °C – 800 °C | 0.014 – 0.05 °C | Fluke 7526A (direct) |
| Temperature (generate & measure)⁵ | −197 °C – 660 °C | 0.019 – 0.05 °C | Hart 1502A w/ 5628 comparison |
| Infrared temperature | 35 °C – 250 °C | 1.5 °C / 1.5% | Fluke 62 MAX+ |
| Relative humidity (generate)⁴ | 10 – 95 %RH | 0.51 %RH | Thunder Scientific 1200 |
| Relative humidity (measure)³ | 0 – 97 %RH | 1.2 – 2.0 %RH | Vaisala HM70 |
Temperature and humidity standards are traceable to NIST through the Fluke 7526A, Hart 1502A reference, and Thunder Scientific / Vaisala humidity standards. Uncertainties are reported at a 95% confidence level (k=2) on each ISO/IEC 17025:2017 certificate.
DC / Low Frequency (Electrical)
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| DC voltage (generate / measure) | 0 mV – 1000 V | scales with range | Fluke 5720A / HP 3458A |
| DC current (generate / measure) | 10 µA – 1000 A | scales with range | Fluke 5720A / 5520A · HP 3458A · LAB100-100 |
| DC resistance (generate) | 0 Ω – 1 TΩ | scales with range | Fluke 5720A / 5520A · Keithley 487 V/I |
| DC resistance (measure) | 0 Ω – 20 GΩ | scales with range | Fluke 8508A (normal + HV) |
| AC voltage (generate) | 0.22 mV – 1100 V (10 Hz–1 MHz) | scales with range | Fluke 5720A / 5725A |
| AC voltage (measure) | 0 mV – 700 V; 0 – 20 kV | scales with range | HP 3458A / Vitrek 4620B |
| Capacitance | 1 pF – 110 mF | scales with range | Fluke 5520A / HP 4284A |
| Inductance | 10 µH – 10 H | scales with range | HP 4284A (100 Hz–100 kHz) |
| Oscilloscope — leveled sine | 5 mVp-p – 5 Vp-p, 0.1 Hz – 3.2 GHz | per Fluke 9530 | Fluke 9500B w/ 9530 |
| Oscilloscope — edge rise/fall | 150 ps / 500 ps | 19 ps / 44 ps | Fluke 9500B (fast edge) |
DC and low-frequency electrical measurements are traceable to NIST through the Fluke 5720A/5520A multifunction calibrators, HP 3458A, Fluke 8508A, and Vitrek 4620B reference standards. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.
RF / Microwave & Time/Frequency
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| RF power (measure)³ | 0.0001 – 26.5 GHz, −20 to +20 dBm | 0.12 – 0.35 dBm | HP 437B w/ HP 8482A (≤18 GHz) / 8485A (≤26.5 GHz) |
| Frequency (reference)⁴ | 0.1 – 10 MHz | per Fluke 910R | Fluke 910R |
| Frequency (generate / measure) | 1 Hz – 26.5 GHz | per timebase | HP 3325 + Fluke 910R timebase |
| Timers / stopwatches | 0.001 s – 99,999.999 s | 0.001 % | VWR 62344-880 (comparison) |
RF power, frequency, and time measurements are traceable to NIST through HP 437B power sensors and the Fluke 910R frequency reference. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.
Chemical
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| pH (generate)⁴ | 10 pH | 0.024 pH | Buffer solutions, direct method |
| pH (generate)⁴ | 7 pH | 0.035 pH | Buffer solutions, direct method |
| pH (generate)⁴ | 4 pH | 0.035 pH | Buffer solutions, direct method |
pH calibrations are traceable to NIST through certified buffer solutions. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.
Pressure & Vacuum
| Parameter | Range | Expanded Uncertainty (k=2) | Reference Standard / Method |
|---|---|---|---|
| Pressure (low / inH₂O) | 0 – 10 inH₂O | 0.031 inH₂O | Fluke 700P01 / 719 |
| Pressure / vacuum (gauge) | −12 – 30 psi | 0.018 psi | Fluke 717-30G |
| Pressure (absolute) | 0 – 30 psia | 0.021 psia | Fluke 700PA5 |
| Pressure (gauge) | 30 – 100 psi | 0.044 psi | Fluke 719-100G |
| Pressure (gauge) | 100 – 300 psi | 0.12 psi | Fluke 719 Pro 300G |
| Pressure (gauge) | 300 – 1,000 psi | 0.51 psi | Fluke 700P08 |
| Pressure (gauge) | 1,000 – 3,000 psi | 2.4 psi | Fluke 700P29 |
| Pressure (gauge) | 3,000 – 5,000 psi | 4.0 psi | Fluke 700P30 |
| Pressure (gauge) | 5,000 – 10,000 psi | 8.1 psi | Fluke 700P31 |
Pressure and vacuum measurements are traceable to NIST through Fluke reference pressure modules. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.
- ¹ The uncertainty covered by the Calibration and Measurement Capability (CMC) is expressed as the expanded uncertainty having a coverage probability of approximately 95 %. It is the smallest measurement uncertainty that a laboratory can achieve within its scope of accreditation when performing calibrations of a best existing device. The measurement uncertainty reported on a calibration certificate may be greater than that provided in the CMC due to the behavior of the calibration item and other factors that may contribute to the uncertainty of a specific calibration.
- ² When uncertainty is stated in relative terms (percent, multiplier, decimal fraction, or scientific notation), it is in relation to instrument reading or instrument output, as appropriate, unless otherwise indicated.
- ³ Capability is suitable for the calibration of measuring devices in the stated ranges.
- ⁴ Capability is suitable for the calibration of devices intended to generate the indicated quantity in the stated ranges.
- ⁵ Also available as site calibration. Note that actual measurement uncertainties achievable at a customer's site can normally be expected to be larger than the uncertainties listed on this Scope of Accreditation.
Measurements are delivered within a 95% confidence level (k=2) with full 'As Found – As Left' data on certificates where applicable.