[Accreditation Scope]DOC FCI-SCOPE-001 · REV C

Published ISO/IEC 17025:2017 scope of accreditation

FCI's accredited calibration and measurement capability (CMC), grouped by discipline. Ranges and expanded uncertainties are reported at a 95% confidence level (k=2). Accredited by IAS under certificate CL-112.

IAS CL-112ISO/IEC 17025:2017NIST TraceableCAGE 5YJ59ANC-OwnedBuy Indian Act43 Years
§2 · 3 parameters

Dimensional

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
Micrometers (direct)0.0625 in – 60 in(70 + 7.5L) µinGage blocks + micrometer standards
Calipers (0–1 in)0 – 1 in280 µinDirect measurement, gage blocks
Calipers (1–60 in)1 in – 60 in320 µinDirect measurement, gage blocks

Dimensional measurements are traceable to the SI metre through NIST-traceable gage-block and micrometer standards. Uncertainties are evaluated per the ISO/IEC Guide 98-3 (GUM) and reported at a 95% confidence level (k=2) on every ISO/IEC 17025:2017 certificate.

§3 · 6 parameters

Mechanical

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
Speed (non-contact, generate)⁴6 – 99,999 rpm0.011 – 0.59 rpmHP 3325B
Scales & balances (linearity)200 mg – 15 kg0.070 mg – 130 mgClass 1 / Class 3 weights
Scales & balances (lb)33 lb – 400 lb0.0058 – 0.055 lbClass 7 weights
Force gauges0.5 – 10 lbf0.0041 – 0.054 lbfClass 7 weights
Force (measure)³0 – 225,000 lbf18 lbf – 1.6e3 lbfInterface 1211NE-10K-B · VLC-120 · Tetrahedron 300K
Hand torque tools (CW + CCW)10 – 3,120 lbf·inper ISO 6789-1Larson UTWCS (ISO 6789-1)

Speed, mass, force, and torque results are traceable to NIST through reference frequency, certified weight sets, reference load cells, and torque standards. All uncertainties are reported at a 95% confidence level (k=2) on each ISO/IEC 17025:2017 certificate.

§4 · 6 parameters

Thermal / Environmental

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
Simulated temperature (thermocouple)−250 °C – 1820 °C0.08 – 0.73 °CFluke 7526A (direct)
Simulated temperature (RTD)−200 °C – 800 °C0.014 – 0.05 °CFluke 7526A (direct)
Temperature (generate & measure)⁵−197 °C – 660 °C0.019 – 0.05 °CHart 1502A w/ 5628 comparison
Infrared temperature35 °C – 250 °C1.5 °C / 1.5%Fluke 62 MAX+
Relative humidity (generate)⁴10 – 95 %RH0.51 %RHThunder Scientific 1200
Relative humidity (measure)³0 – 97 %RH1.2 – 2.0 %RHVaisala HM70

Temperature and humidity standards are traceable to NIST through the Fluke 7526A, Hart 1502A reference, and Thunder Scientific / Vaisala humidity standards. Uncertainties are reported at a 95% confidence level (k=2) on each ISO/IEC 17025:2017 certificate.

§5 · 10 parameters

DC / Low Frequency (Electrical)

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
DC voltage (generate / measure)0 mV – 1000 Vscales with rangeFluke 5720A / HP 3458A
DC current (generate / measure)10 µA – 1000 Ascales with rangeFluke 5720A / 5520A · HP 3458A · LAB100-100
DC resistance (generate)0 Ω – 1 TΩscales with rangeFluke 5720A / 5520A · Keithley 487 V/I
DC resistance (measure)0 Ω – 20 GΩscales with rangeFluke 8508A (normal + HV)
AC voltage (generate)0.22 mV – 1100 V (10 Hz–1 MHz)scales with rangeFluke 5720A / 5725A
AC voltage (measure)0 mV – 700 V; 0 – 20 kVscales with rangeHP 3458A / Vitrek 4620B
Capacitance1 pF – 110 mFscales with rangeFluke 5520A / HP 4284A
Inductance10 µH – 10 Hscales with rangeHP 4284A (100 Hz–100 kHz)
Oscilloscope — leveled sine5 mVp-p – 5 Vp-p, 0.1 Hz – 3.2 GHzper Fluke 9530Fluke 9500B w/ 9530
Oscilloscope — edge rise/fall150 ps / 500 ps19 ps / 44 psFluke 9500B (fast edge)

DC and low-frequency electrical measurements are traceable to NIST through the Fluke 5720A/5520A multifunction calibrators, HP 3458A, Fluke 8508A, and Vitrek 4620B reference standards. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.

§6 · 4 parameters

RF / Microwave & Time/Frequency

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
RF power (measure)³0.0001 – 26.5 GHz, −20 to +20 dBm0.12 – 0.35 dBmHP 437B w/ HP 8482A (≤18 GHz) / 8485A (≤26.5 GHz)
Frequency (reference)⁴0.1 – 10 MHzper Fluke 910RFluke 910R
Frequency (generate / measure)1 Hz – 26.5 GHzper timebaseHP 3325 + Fluke 910R timebase
Timers / stopwatches0.001 s – 99,999.999 s0.001 %VWR 62344-880 (comparison)

RF power, frequency, and time measurements are traceable to NIST through HP 437B power sensors and the Fluke 910R frequency reference. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.

§7 · 3 parameters

Chemical

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
pH (generate)⁴10 pH0.024 pHBuffer solutions, direct method
pH (generate)⁴7 pH0.035 pHBuffer solutions, direct method
pH (generate)⁴4 pH0.035 pHBuffer solutions, direct method

pH calibrations are traceable to NIST through certified buffer solutions. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.

§8 · 9 parameters

Pressure & Vacuum

Detail →
ParameterRangeExpanded Uncertainty (k=2)Reference Standard / Method
Pressure (low / inH₂O)0 – 10 inH₂O0.031 inH₂OFluke 700P01 / 719
Pressure / vacuum (gauge)−12 – 30 psi0.018 psiFluke 717-30G
Pressure (absolute)0 – 30 psia0.021 psiaFluke 700PA5
Pressure (gauge)30 – 100 psi0.044 psiFluke 719-100G
Pressure (gauge)100 – 300 psi0.12 psiFluke 719 Pro 300G
Pressure (gauge)300 – 1,000 psi0.51 psiFluke 700P08
Pressure (gauge)1,000 – 3,000 psi2.4 psiFluke 700P29
Pressure (gauge)3,000 – 5,000 psi4.0 psiFluke 700P30
Pressure (gauge)5,000 – 10,000 psi8.1 psiFluke 700P31

Pressure and vacuum measurements are traceable to NIST through Fluke reference pressure modules. Uncertainties are reported at 95% confidence (k=2) on each ISO/IEC 17025:2017 certificate.

NOTEThis summary reflects representative accredited parameters. The complete, controlled scope document is available on request and from the IAS directory under certificate CL-112. CMC values may be further constrained by instrument and method.
CMC Footnotes
  1. ¹ The uncertainty covered by the Calibration and Measurement Capability (CMC) is expressed as the expanded uncertainty having a coverage probability of approximately 95 %. It is the smallest measurement uncertainty that a laboratory can achieve within its scope of accreditation when performing calibrations of a best existing device. The measurement uncertainty reported on a calibration certificate may be greater than that provided in the CMC due to the behavior of the calibration item and other factors that may contribute to the uncertainty of a specific calibration.
  2. ² When uncertainty is stated in relative terms (percent, multiplier, decimal fraction, or scientific notation), it is in relation to instrument reading or instrument output, as appropriate, unless otherwise indicated.
  3. ³ Capability is suitable for the calibration of measuring devices in the stated ranges.
  4. ⁴ Capability is suitable for the calibration of devices intended to generate the indicated quantity in the stated ranges.
  5. ⁵ Also available as site calibration. Note that actual measurement uncertainties achievable at a customer's site can normally be expected to be larger than the uncertainties listed on this Scope of Accreditation.

Measurements are delivered within a 95% confidence level (k=2) with full 'As Found – As Left' data on certificates where applicable.